Analysis of Interpenetrating Polymer Networks via Quartz Crystal Microbalance with Dissipation Monitoring
A quartz crystal microbalance with dissipation monitoring (QCM-D) was used to assess the physical properties of interpenetrating polymer networks (IPNs) through swelling experiments in ambient humidity and in phosphate-buffered saline (PBS), pH 7.4. The IPNs, based on acrylamide (AAm) and poly(ethylene glycol) (pEG), swell from thin, rigid films when dry (16.7 ± 5.2 nm on Si/SiO2) to expanded, viscoelastic films when hydrated (107 ± 24.2 nm on Si/SiO2). The dry IPNs could be analyzed using the Sauerbrey relationship, but for the hydrated films it was necessary to interpret QCM-D data with a Kelvin-Voigt viscoelastic model. A complex modulus lG*l of 116 ± 38.1 kPa for the swollen IPN surface on Si/SiO2 was defined by the model. The QCM-D was also employed to quantify the adsorption of human fibrinogen, a protein important in thrombus formation, onto the IPNs. Fibrinogen adsorption studies demonstrated the sensitivity of the QCM-D, as well as confirmed the nonfouling nature of the IPN surface, where less than 5 ng/cm2 of fibrinogen was adsorbed.